AH Series – Hydrophone Preamplifier and Accessories

  • Low-noise preamplifiers for precise hydrophone measurements
  • Frequency range from 50 kHz to 100 MHz
  • Fully submersible designs
  • AH-2010 with 20 dB gain
  • AH-2020 switchable between 0 and 20 dB
  • Accessories such as attenuators, adapters, cables, and mounts
RK

Ralf Kott

Raman and NIR Spectroscopy, Measurement Technology


Download data sheet (PDF)

AH Series – Preamplifiers and accessories for Onda hydrophones

The Onda AH-Series includes low-noise preamplifiers and accessories for reliable signal conditioning in precise hydrophone measurements. The preamplifiers are electronically and mechanically matched to Onda hydrophones, combining a wide frequency range with low self-noise and high stability.

With various amplifier configurations as well as attenuators, adapters, cables, and brackets, the measurement chain can be specifically adapted to the respective ultrasonic application.

Broadband signal conditioning up to 100 MHz

The preamplifiers AH-2010 and AH-2020 offer a frequency range of 50 kHz to 100 MHz and are therefore suitable for broadband ultrasonic measurements.

A flat frequency response and low self-noise enable reliable transmission of the signals generated by the hydrophone. The output impedance is 50 Ω, allowing direct connection to typical oscilloscopes and measuring instruments.

AH-2010 and AH-2020

The AH-2010 operates with a fixed voltage gain of 20 dB and offers a simple solution for measurement tasks with constant signal amplification requirements.

The AH-2020 expands this concept with a switchable gain of 0 or 20 dB. Switching can be done manually or via a TTL signal. This allows the preamplifier to be flexibly adapted to different signal levels within a measurement setup.

Submersible design for compact measurement setups

The AH preamplifiers are fully submersible and can be connected to the hydrophone directly inside the water tank. As a result, the first amplifier stage is located close to the sensor, and only a coaxial cable needs to be routed out of the tank to the measuring device.

The waterproof connection to Onda hydrophones enables compact and low-noise measurement setups, particularly in automated hydrophone scanning systems.

Accessories for the complete hydrophone measurement chain

In addition to the preamplifiers, Onda offers comprehensive accessories for configuring the measurement chain. These include, among others Attenuators, adapters, coaxial cables, and hydrophone mounts.

The ATH-2000 Attenuator can be used, for example, to extend the usable dynamic range of the measurement chain and prevent overloading of the preamplifier at high signal levels.

calibrated measuring chain

The preamplifiers can be calibrated together with the hydrophone used. This is particularly relevant for high-frequency measurements: Onda points out that above 20 MHz a calibration of the hydrophone together with the preamplifier actually used is required, since cable and impedance effects are having an increasing influence on the measurement.

This allows the entire signal chain comprising the hydrophone, preamplifier, and cabling to be specifically configured for quantitative ultrasonic measurements.

Typical applications

  • Hydrophone measurements in water tanks
  • Characterization of ultrasonic fields
  • Medical ultrasound technology
  • Development and testing of ultrasonic transducers
  • Automated hydrophone scanning systems
  • broadband ultrasonic measurements
  • Research and Development
  • Quality assurance and calibration

Benefits of the AH series

  • Frequency range from 50 kHz to 100 MHz
  • Low self-noise
  • Flat frequency response
  • Fully submersible design
  • Waterproof connection to Onda hydrophones
  • AH-2010 with 20 dB gain
  • AH-2020 with switchable 0/20 dB
  • Just one cable from the water tank
  • Calibration available
  • Extensive accessories for complete measuring chains

Preamplifiers and accessories from Onda

The AH Series and the matching Onda accessories enable the setup of a matched measurement chain from the hydrophone to the measuring instrument. Various preamplifiers, attenuators, adapters, and cables allow adaptation to different signal levels, frequency ranges, and measurement geometries.

Soliton assists you in selecting and configuring hydrophones, preamplifiers, cabling, and other accessories for your ultrasonic measurement.

Specifications

AH-2010-100 AH-2020-100
Bandwidth 50 kHz - 100 MHz
Voltage Gain 20 dB -Hand or TTL switchable 0 or 20 dB
Input Impedance 1 MΩ in parallel with 6 pF 1 MΩ in parallel with 7 pF
Output Impedance 50 Ω
Max output into 50 Ω 4 Vpp below 50 MHz
1 Vpp above 50 MHz
1.2 Vpp
Power consumption (max) 2 W 2.5 W
Noise Level 20 dB: 360 μVrms 20 dB: 400 μVrms
0 dB: 350 μVrms
Calibration Included